single channel Pipette
Lightweight, ergonomic, low force design
Digital display clearly reads volume setting
The pipettes cover volume range of 0.1μl to 10ml
Easy to calibrate and maintain with tool supplied
Design helps avoid repetitive strain injuries
Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
The low part is available for autoclaving
Lightweight, ergonomic, low force design
Digital display clearly reads volume setting
The pipettes cover volume range of 0.1μl to 10ml
Easy to calibrate and maintain with tool supplied
Design helps avoid repetitive strain injuries
Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
The low part is available for autoclaving
ROOM.106, BUILDING 10, NO.155, SHANGYUAN STREET, JIANGNING DISTRICT, NANJING, CHINA
Mr Aaron Dong; Ms Alice
+86-025-58095296
+86-13675117634